Get Quick Quote From Supplier
1.The unit must at least be an electrically powered, dry warmer, to warm blood stored under refrigeration and other infused fluids.2.The unit shall be factory fitted with an illuminated on/off switch and safety over load fuse on rear panel. 3.The unit must at least be able to warm the fluid to the desired temperature within a minimum of 30 seconds.4.The unit must at least be transportable.5.The unit shall not weigh more than 90kg.6.The unit interior shall be manufactured from 304L stainless steel.7.The unit shall be factory fitted with an epoxy coated electro-galvanised outer case.8.The unit shall be factory fitted with a high temperature glass wool isolation.9.The unit shall be factory fitted with anti-tilt shelf runners with stainless steel wire shelves.10.The unit be factory fitted with forced air circulation and internal baffles for temperature stability.11.The unit temperature range shall at least be between -5u02daC above ambient to 99.9u02daC.12.The unit shall be factory fitted with Heater u2013 onu2019u2019 indicator light.13.The unit shall be factory fitted with at least a two (2) meter extension lead and a 16A 3-pin plug.14.The unit shall be factory fitted with a variable over-temperature cut-out with indicator light.